AUTO SALINITY TEMPERATURE METER NDS06

 

is an ideal instrument for measurement of Salinity of aqueous solutions. They use the latest microcontroller technology to measure the Salinity in three ranges from 2 ppt to 50 ppt. The appropriate range is automatically selected by the instrument. The Resolution is 0.001 ppt to 1 ppm in the lowest range. The cell constant is flashed on a digital display and can be adjusted from the front panel. Facility is also provided for automatic measurement of cell constant with a known standard solution.

Features :

  • Microcontroller Based
  • Auto Ranging Facility
  • Auto Cell Constant Measurement
  • 16x2 LCD Display

This instrument is ideal for monitoring salt contents in water and soluble salts in soils. This is extremely useful instrument for agriculture and soil analysis labs, swimming pools, fertilizer plants, mineral water plants, petroleum refineries, textile plants etc.

Product Details:

Parameter Salinity Temperature
Range 0 to 200 ppm, 0 to 20.00 ppt, 0 to 50.0 ppt 0°C to 100°C
Resolution 0.1ppt 0.1°C
Accuracy ± 0.5% of range±1 digit ± 0.1°C + 1 digit of range
Range Selection Automatic
Temp .Compensation Auto
Salinity Factor 0.75
Cell Constant adjustment Digital : 0.05 to 2.000 By default 1.000
Reference Temperature 20°C& 25°C
Measuring frequency 1000Hz
Mode of calibration With Standard Solutions
Functions Salinity & Temperature
Sensor Temperature probe
Display LCD
Power Supply 230 V ±10%, 50 Hz AC
Dimensions & Weight (L x B x H) 195 x 225 x 65 mm: 3 kgs
Accessories Salinity Cell, Dust Cover, Instruction manual, Standard solutions, Temperature probe.